122 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 20: Saving image data
Saving image data
You can save your image data in one of two ways:
Save your data as a multiple frame animation of your image frames. This
ultrasound image is called a cine loop.
Save your data as a single frame ultrasound image called an image frame.
In this chapter
Saving a cine loop (multiple-frame animation)...............................................................122
Saving an image frame .......................................................................................................123
Saving a cine loop (multiple-frame animation)
A cine loop is a multiple -frame animation of your image frames. You can save
your image data as a cine loop in every image Mode other than 3D-Mode.
B-Mode based cine loops are measured by number of frames. M-Mode and PW
Doppler Mode cine loops are measured in seconds.
How cine loops work
While you acquire data, the system's playback memory holds your most recent
image data in a buffer. The size of the buffer is determined by the Cine Loop Size
preference you specify in the Preferences window on the General tab.
When you save your image as a cine loop, the system saves this buffered data as
an image. The buffer saves the latest acquired data.
f To review your cine loop content before you save it:
1. Press
Scan/Freeze
.
2. Use the
Cine Loop Review
dial to review the current, but unsaved, cine loop
frames.
3. If you don't want to save the content, press
Scan/Freeze
again and continue to
acquire new image data.
f To save your image as a cine loop:
1. Press
Scan/Freeze
to stop acquiring data.