194 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 33: Acquiring B-Mode images
Control panel controls for B-Mode
When you are acquiring B-Mode image data, these are the controls you use to
optimize the image you see on the screen.
Image Width
Adjusts the physical width of the area the transducer is imaging. Push up to increase the
width. Pull down to decrease the width.
Tip: The closer you can reasonably narrow the width of your image around your target
structure, the higher the system sets the acquisition frame rate. This is especially helpful
when you are studying cardiac tissue movement.
Display Map
Cycles you through a predefined set of optimization maps that you can apply either
while you are acquiring or reviewing image data.
Push up or pull down to cycle through the available maps for the active imaging mode.