308 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 43: Acquiring Color Doppler Mode images
Area Description
Region of interest color box overlay. The system applies the Color Doppler Mode based
colors only to the image data within this box.
Vascular flow moving toward the transducer. Displayed in red colors.
Vascular flow moving away from the transducer. Displayed in blue colors.
Color and velocity scale. The right column of the scale is the color scale. It follows the
acronym BART color principle for Doppler (Blue=Away from, Red=Toward) positive
vascular flows are indicated by colors in the red range, negative flows are in the blue
range, and velocities for each direction increase from dark to light. The velocity range of
the scale changes when you change the signal velocity or frequency.
The left column of the scale is the standard gray scale that appears for all B-Mode based
images.
Priority indicator. Tracks the priority level when you adjust the
Priority
control. This
control adjusts the priority relationship between the overlay data and the background B-
Mode data so you can eliminate false readings. For more information see Priority (page
415).
Physiological data trace window. Displays your animal's heart rate, temperature,
respiration rate and blood pressure data. During data acquisition this information comes
from the Advanced Physiological Monitoring Unit connected to the Vevo Imaging Station.
Cine loop range control. Displays the length of the cine loop range. The triangular white
marker identifies the individual frame number within the cine loop. You can drag the left
and right vertical markers to display only the image frames in that range.
Live physiological display. If the animal is connected to the physiology controller, data
appears here in real time during image acquisition and can display the numeric values of
the animal's heart rate, respiration rate, blood pressure and body temperature. This area
also displays the image data storage capacity progress bar so you can see when you
should start to back up your image data to free up space on the system. Live physiological
data is only active when you enable the inputs in the General tab of the Preferences
window.
Screen keys display
Displays the updated parameter and system information when you make adjustments
on the control panel.
Displays control options in the mode that you apply during image acquisition when
you press the
Screen Keys
dial.
In Color Doppler Mode, press the dial to cycle through three image states: Color box
overlay + B-Mode, B-Mode only, Color box only.