206 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 34: Analyzing B-Mode images
In this example, six wall points have been added to the trace curve
Right-click the final point on the contour to complete the measurement. If you
selected the Show Values and Labels option in the Measurements tab of the
Preferences window, the system displays the measurement value and editable
label for the measurement B-Mode LV Area #, where # is a sequential
number.
6. If you want to rename the label and you have selected Show Values and
Labels in the Measurement tab of the Preferences window, type a new name
while the label text is selected, and then click outside the label to commit the
label.
7. If you have selected Show Values and Labels in the Measurement tab of the
Preferences window and you want to move the measurement or move the
label, select either item and then drag and drop it.
8. Move to another frame in the cine loop and place another LV area long axis
measurement.
9. Right-click the contour and select Replicate Forward or Replicate Reverse
with additional options to define how many cycles: either 2 or 3. The system
automatically traces the wall forward or backward through the frames.
Frame 1: traced manually The system traces the remaining frames automatically
10. Modify the contour or points on the contour if required and then select
Replicate Forward again to complete the automatic wall trace.
11. Press
Cine Store
to save the cine loop.
When you play the cine loop, the system displays the contour that represents the
systolic LV in green, and the diastolic LV in red.
Next step
Reporting your analysis results (page 184)
Related information
Analyzing image data (page 156)