Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 75
Chapter 10: Setting the General tab preferences
Preference Description
Sequence
Destroy
Position
From the drop-down list select the moment in the pre-triggered cine loop when the
system begins the destruction burst. The value is set as a percentage. For example,
if your cine loop is set to 200 frames and you set the value to 25%, the system will
run the destroy burst at frame 50.
3. Click OK.
Physiological Enable preferences
Use the Physiological Enable options to globally enable or disable the system's
ability to save the Respiration, Blood Pressure and Temperature physiological
signal inputs along with the ultrasound image.
How physiological data inputs work
The system receives the physiological signal inputs from the Advanced
Physiological Monitoring Unit through the Physio Data port on the rear panel of
the cart.
Enabling or disabling an input determines whether or not you can work with it in
other workspaces in the system.
When you select an input, you can control whether or not to control the display of
the real-time physiological data in two places:
The Physiological Live Display section of the General tab in the Preferences
window.
For example, as illustrated below, if you select the Respiration check box but
clear the Blood Pressure and Temperature check boxes, you will only see the
Respiration display control check box. (Note: The ECG signal input cannot be
disabled, so you will always be able to control whether or not to display it.)
In this example, you would only be able to show or hide the Respiration data
in the physiological live display strip at the bottom right corner of the screen.
The Physiological Display section in the Physio Options left panel display in
a mode window.