266 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 39: Analyzing PW Doppler Mode and PW Tissue Doppler Mode images
Related information
Analyzing image data (page 156)
Applying automatic traces to the frequency waveform (page 267)
Heart rate measurement
Use the heart rate measurement tool for measuring the average heart rate (in
BPM) of an animal by measuring the distance over time between the displayed
cardiac cycles.
f To place a heart rate measurement:
1. Click the heart rate measurement button .
2. Click on your image to place the initial caliper at a specific point in the cardiac
cycle.
3. Trackball to the same location on the next cardiac cycle and click to place the
next caliper.
4. Continue placing calipers on the cardiac cycles and then right-click on the last
heart beat of the sequence to place your final caliper.
5. If you need to move the entire measurement, click on a measurement line and
then drag and drop it.
Related information
Complete procedure for adding a measurement (page 166)
Single point measurement
Use the linear distance measurement tool to place a caliper dot on the image. A
single point measurement records the following properties of the dot:
Cine loop time point measured in ms
Doppler frequency measured in KHz
Velocity measured in mm/s
f To place a single point measurement:
1. While you view a saved image from the Study Browser or an image that is
acquired but not stored during an image acquisition session, press
Measure
and toggle to view the measurement tools panel.
2. Click the single point measurement button . The system highlights the
button until you complete your measurement.