264 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 39: Analyzing PW Doppler Mode and PW Tissue Doppler Mode images
3. Click on your image to place the initial caliper at a specific point on the
waveform.
4. Trackball along the contour of the waveform. The system automatically places
points at the spacing density that you specify in the Auto Point Spacing
section of the Measurement tab in the Preferences window.
5. Right-click to place your final caliper at the end of the last cardiac cycle. If you
selected the Show Values and Labels option in the Measurements tab of the
Preferences window, the system displays the measurement value and editable
label for the measurement VTI #, where # is a sequential number.
The system constrains your VTI measurement to either the top or bottom side of the
baseline, depending on the placement of the initial caliper. The system prevents your
trace from crossing the baseline.
6. If you want to position points more accurately over regions where the signal
changes rapidly, drag them into position.
7. If you want to rename the label and you have selected Show Values and
Labels in the Measurement tab of the Preferences window, type a new name
while the label text is selected, and then click outside the label to commit the
label.
8. If you have selected Show Values and Labels in the Measurement tab of the
Preferences window and you want to move the measurement or move the
label, select either item and then drag and drop it.
Next step
Reporting your analysis results (page 184)
Related information
Analyzing image data (page 156)