Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 173
Chapter 30: Working with measurements
8. In the protocols list select the protocol measurement you want to work with
and then add the measurement on the image.
M-Mode measurement chains
In M-Mode, you can complete the following sequenced measurements in
automatic chains, as shown in the following diagram:
M-Mode image displaying the measurement chains beginning with RVID;d and IVS;s
In the sequence of chained measurements, the final caliper of the first
measurement in the chain automatically becomes the first caliper of the second
measurement. This linking continues for the remainder of the caliper points.
The labeling for all measurements occur at the same time and only when you add
the last caliper of the final measurement in the chain. The image is stored as each
of the measurements is completed.
f To complete an M-Mode chained measurement:
1. In the measurement packages list select Cardiac Package.
2. In the protocols list, click the protocol and then click the first measurement in
the chain. For example, click PSLAX > RVID;d.
3. Click the top point of the first measurement of the chain and move the cursor
toward the bottom point. For example, click the top point of the RVID;d
measurement.
The system displays and labels the measurement if the Show Values and
Labels option is selected in the Measurement tab of the Preferences window.
4. Click the bottom point of the first measurement. The system commits the
measurement value for the first measurement and stores the image.