Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 241
Chapter 36: Analyzing M-Mode images
Adding M-Mode measurement chains
In M-Mode, the most precise way to create diastole and systole measurement sets
is to stack your measurements.
To automate this procedure the system automatically links the following
measurements into chained sequences. For example, if you select the Cardiac
Package and then select the SAX (short axis) protocol, you can create the
following diastole and systole measurement chains:
Diastole measurement chains
IVS --> LVID --> LVPW
LVAW --> LVID --> LVPW
Systole measurement chains
IVS --> LVID --> LVPW
LVAW --> LVID --> LVPW
f To add a complete chained measurement:
1. In the protocol measurements list, click the first measurement in the chain.
2. Click the top point of the first measurement of the chain and move the cursor
toward the bottom point.
The system labels the measurement and displays the measurement value
dynamically as the cursor is moved toward the bottom point.
3. Click the bottom point of the first measurement. The system commits the
measurement value for the first measurement.
This bottom point of the first measurement automatically becomes the top
point of the second measurement in the chain.
4. Click the bottom point of the second measurement. The system measures and
labels the second measurement.
5. Click the remaining bottom points of the next measurements in the chain. The
system measures and labels each measurement until the final measurement is
completed.
f To add individual measurements from a chain:
1. In the protocol measurements list, click any one of the measurements in the
chain.
2. Press
ESC
to cancel the chain but keep the completed measurements.